
KUDUwave Portable Audiometers — Diagnostic Testing Without a Fixed Booth
Boothless audiometers substitute a fixed sound booth with sound-attenuating earcups and continuous ambient-noise monitoring, so a compliant hearing test can be administered inside a facility rather than only in a dedicated clinic. In a correctional setting, that mobility avoids the security and transport burden of moving individuals off-site for a routine screening.
Audiometric testing generally references ANSI/ASA S3.6, the standard that specifies audiometer performance, and OSHA's noise standard requires that hearing-conservation tests be conducted in a suitably quiet environment. Integrated noise monitoring is designed to document that ambient conditions were within acceptable limits for each test, supporting the recordkeeping a hearing conservation program needs.
Any diagnostic device still has to be calibrated and maintained to remain accurate; audiometers are typically calibrated on a defined schedule. Claims about a specific device delivering booth-equivalent results should be confirmed against the manufacturer's own validation and regulatory clearance documentation.
Sources: OSHA 29 CFR 1910.95; ASHA































